• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON CARIBE LTD. BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON CARIBE LTD. BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING Back to Search Results
Catalog Number 442023
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/22/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter facility name: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported that while using bd bactec¿ plus aerobic/f culture vials (plastic), there was found to be bactieral contamination in the vials.This is a report of 7 occurrences.No false results were reported out to the clinician or the patients.The following information was provided by the initial reporter: 7 bactec vials turned positive for bacillus cereus in one day.Four different lot numbers, four different patients, two different ward stations.
 
Manufacturer Narrative
H6: investigation summary: catalog: 442023.Batch no.: 2041398.Customer reported a contamination issue while using bactec product.Neither photos nor returned good samples were received.Bd was unable to reproduce the customer¿s experience with bactec product.Retention samples were visually inspected, tested for viable contamination by sub-culturing on tsa, chocolate, sabouraud and schaedler agars plates, voltage output and gram stain.All results were satisfactory.Batch history record review did not identify any evidence for which the customer submitted the complaint.A complaint history review was conducted and only the current complaint was found relating to the incident lot number and the ¿as reported¿ defect code.Complaint is unconfirmed based on retention samples and batch history record review results.Product insert warnings and precautions sections states that prior to use, each vial should be examined for evidence of contamination such as cloudiness, bulging or depresses septum, or leakage.Vials showing evidence of contamination should not be used.Also prior to use, the user should examine the vial for evidence of damage or deterioration.Vials displaying turbidity, contamination, or discoloration (darkening) should not be used.No corrective actions were required.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigational process.
 
Event Description
It was reported that while using bd bactec¿ plus aerobic/f culture vials (plastic), there was found to be bacterial contamination in the vials.This is a report of 7 occurrences.No false results were reported out to the clinician or the patients.The following information was provided by the initial reporter: 7 bactec vials turned positive for bacillus cereus in one day.Four different lot numbers, four different patients, two different ward stations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)
Type of Device
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer (Section G)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15171819
MDR Text Key304967489
Report Number2647876-2022-00177
Device Sequence Number1
Product Code MDB
UDI-Device Identifier00382904420239
UDI-Public382904420239
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K113558
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2022
Device Catalogue Number442023
Device Lot Number2041398
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-