• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RENEGADE HI-FLO; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RENEGADE HI-FLO; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 29420
Device Problem Difficult to Open or Remove Packaging Material (2922)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2022
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 19jul2022.It was reported that the microcatheter would not release from the hoop.A 105/10 renegade hi-flo microcatheter was selected for use.During unpacking, it was noted that the microcatheter would not release from the hoop.After flushing the inside and the outside of the microcatheter, it would not release.The procedure was completed with another of the same device.No patient complications were reported.However, device analysis revealed microcatheter fracture located at 22.7cm from the hub.
 
Manufacturer Narrative
Device evaluated by mfr: the renegade hi flo was returned to boston scientific for analysis.The device was inspected for any damage or irregularities.The renegade hi flo was received completely removed from the shipping hoop.A fracture 22.7 cm from the hub was visible.The strain relief was noted to be on the distal portion of the fracture and not on the hub.Signs of stretching were noted 16.4 cm from the hub.Of note, signs of saline were seen on the distal portion of the catheter, but the proximal end of the catheter showed no signs of saline.Because there was no evidence of any product quality deficiencies, it was considered likely that the fractures and stretching attributable to handling issues.This concludes the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEGADE HI-FLO
Type of Device
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15193535
MDR Text Key304907068
Report Number2134265-2022-08402
Device Sequence Number1
Product Code KRA
UDI-Device Identifier08714729280590
UDI-Public08714729280590
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140329
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number29420
Device Catalogue Number29420
Device Lot Number0028570196
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-