LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2022
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message associated to the stirrer motor on the patient side during procedure.There was no patient injury.
 
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in san diego, california.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to a blocked stirrer motor.The part was replaced and the problem could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
H.10: the unit is installed since 2010 and a review of the complaints database revealed that a similar event on this heater cooler was submitted in 2021 (manufacturer report number 9611109-2021-00600).Based on service activity findings and device service history review, the most likely cause of the reported issue is associated to motor bearing damaged by the corrosion, which consequently does not allow the motor shaft to rotate.Possible root causes can be related to environmental/use conditions and cleaning practice in place at the customer site cannot be excluded as contributing factor to the reported failure of the stirrer motor.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15198573
MDR Text Key298045782
Report Number9611109-2022-00407
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-