MAUDE Adverse Event Report: CORDIS CORPORATION SABER RX 5MM30CM 155; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION SABER RX 5MM30CM 155; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number 51005030L
Device Problem Burst Container or Vessel (1074)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2022
Event Type  malfunction  
Manufacturer Narrative
As reported, during the second inflation of a 5mm x 30cm 155cm saberx percutaneous transluminal angioplasty (pta) balloon catheter, the balloon ruptured at eight atmospheres (atm).As a result, a new saberx balloon catheter of an unknown size was used in place of the ruptured 5mm x 30cm saberx pta.There was no reported injury to the patient.This was during an endovascular therapy (evt) procedure to treat a lesion with a 90% stenosis, moderate calcification, and mild vessel tortuosity.The lesion was not a chronic total occlusion (cto).The device was stored and prepped per the instructions for use (ifu) and maintained negative pressure during preparation.There was no difficulty removing the device from the hoop, removing the protective balloon cover, the stylet, or any of the sterile packaging components and the device was not kinked or damaged in any way prior to being inserted into the patient.There was no resistance or friction experienced during insertion of the device.The device was able to track toward the lesion and cross the lesion without difficulty.The device was never put into an acute bend and did not kink during its use.The device was able to be removed easily from the patient and remained in one piece during its removal.The product was not returned for analysis as it was discarded.A product history record (phr) review of lot 82221002 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon- burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics of moderate calcification with stenosis and mild tortuosity may have contributed to the reported event.Calcification is known to damage balloon material; it is likely this occurred when attempting to cross the stenosed/calcified lesion.However, without the return of the device for analysis, it is difficult to draw a clinical conclusion between the device and the event reported.According to the warnings in the safety information in the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Event Description
As reported, during the second inflation of a 5mm x 30cm 155cm saberx percutaneous transluminal angioplasty (pta) balloon catheter, the balloon ruptured at 8 atmospheres (atm).As a result, a new saberx balloon catheter of an unknown size was used in place of the ruptured 5mm x 30cm saberx pta.There was no reported injury to the patient.This was during an endovascular therapy (evt) procedure to treat a lesion with a 90% stenosis, moderate calcification, and mild vessel tortuosity.The lesion was not a chronic total occlusion (cto).The device was stored and prepped per the instructions for use (ifu) and maintained negative pressure during preparation.There was no difficulty removing the device from the hoop, removing the protective balloon cover, the stylet, or any of the sterile packaging components and the device was not kinked or damaged in any way prior to being inserted into the patient.There was no resistance or friction experienced during insertion of the device.The device was able to track toward the lesion and cross the lesion without difficulty.The device was never put into an acute bend and did not kink during its use.The device was able to e removed easily from the patient and remained in one piece during its removal.The device was discarded and will not be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER RX 5MM30CM 155
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key15199748
MDR Text Key304594003
Report Number9616099-2022-05895
Device Sequence Number1
Product Code LIT
UDI-Device Identifier20705032076033
UDI-Public(01)20705032076033(17)240430(10)82221002
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number51005030L
Device Lot Number82221002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/20/2022
Date Device Manufactured05/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SABERX PTA (SIZE UNKNOWN)
-
-