SMITH & NEPHEW, INC. FOOTPRINT ULTRA PK SUTURE ANCHOR 4.5; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Model Number 72202901 |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/20/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a shoulder cuff repair surgery, when the package was opened, the head of the 4.5mm footprint ultra pk suture anchor was split and could not be used.The procedure was successfully completed using a back-up device, but it is unknown if there was a surgical delay due to the reported event.No further complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Manufacturer Narrative
|
H10: h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A review of the customer provided image found the anchor attached to the inserter with the stay suture unwrapped from the handle.No split or fracture of the anchor can be seen in the image.The complaint was not confirmed.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
|
Manufacturer Narrative
|
H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A visual inspection of the returned device found that it is not in its original packaging.The insertion device, anchor and some suture material were returned.The anchor is fractured at the proximal end and there is biological debris in the anchor portion that was retained on the insertion device and on the suture material.A review of the customer provided image found the anchor attached to the inserter with the stay suture unwrapped from the handle.No split or fracture of the anchor can be seen in the image.Based on the condition of the product material found during visual inspection, additional material testing is not required.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
|