Model Number 445870
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2022
Event Type  malfunction  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.
 
Event Description
It was reported that while using bd bactec¿ mgit¿ 960 system a false negative result was obtained.The following information was provided by initial reporter bottle was flagged as negative but analysis of the culture sediment in the bottom of the tube indicated growth.
 
Event Description
It was reported that while using bd bactec¿ mgit¿ 960 system a false negative result was obtained.The following information was provided by initial reporter bottle was flagged as negative but analysis of the culture sediment in the bottom of the tube indicated growth.
 
Manufacturer Narrative
H.6 investigation summary: this complaint alleges a failure on a mgit 960 instrument (p/n 445870, (b)(6)).The customer called in to report a false negative result on their instrument.No instrument errors or issues were noted.Bd technical service made four attempts to follow up with the customer with no response.No additional information is available.No samples or parts were returned for this complaint and thus, returned sample analysis cannot be completed.Review of device history record for instrument serial number, (b)(6) is not required because this complaint does not allege an early life failure or failure at installation and has changed configuration since release from manufacturing due to service repairs/pms.Service history review was performed for the instrument and no additional work orders were observed for the complaint failure mode reported.Root cause is unknown at this time.This is an unconfirmed failure of a bd product.Complaint history for results was reviewed for the month of august.The upper control limit was not breached, and trends were not identified associated with this defect.No new trends, risks, or hazards were identified as a result of this complaint.The issue in this complaint does not require the initiation of a corrective and preventative action (capa).This complaint will be included in the periodic trend review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ MGIT¿ 960 SYSTEM
Type of Device
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15203724
MDR Text Key304975503
Report Number1119779-2022-01090
Device Sequence Number1
Product Code MDB
UDI-Device Identifier00382904458706
UDI-Public00382904458706
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K974883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number445870
Device Catalogue Number445870
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/27/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-