ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PRODIGY MRI IPG, 16CH, RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3772
Device Problem Wireless Communication Problem (3283)
Patient Problem Inadequate Pain Relief (2388)
Event Date 07/21/2022
Event Type  Injury  
Manufacturer Narrative
(b)(6).Date of event is estimated.
 
Event Description
It was reported the patient lost therapy and could no longer charge the ipg.In turn, the ipg was deemed inoperable.As a result, the ipg was explanted and replaced to address the issue.
 
Manufacturer Narrative
The complaint for inoperable ipg was confirmed.The ipg was able to communicate with a patient programmer.However, the ipg would not charge with the lab charging system.The ipg was functionally tested using autotest.The autotest test failed due to high current.Troubleshooting revealed that the inability to charge and high current was isolated to c52, a capacitor in the low battery/current measure circuit.Electrical testing confirmed capacitor c52 was degraded and resulted in the inability to charge the ipg.With capacitor c52 taken out of circuit, the ipg was able to charge normally.The ipg was functionally tested again using autotest and passed all test.Additionally, the device history record of the device was reviewed to confirm that all manufacturing steps were completed and there were no non-conformances noted that could have contributed to this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODIGY MRI IPG, 16CH, RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15210620
MDR Text Key297705967
Report Number3006705815-2022-16073
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017093
UDI-Public05415067017093
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/16/2021
Device Model Number3772
Device Catalogue Number3772
Device Lot NumberA000079010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-