Catalog Number 114604
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(6). The actual device was not available; however, three photographs of the sample were provided for evaluation. Visual inspection of the provided pictures showed in picture one (1) the product clamped in the mechanical device. Water stains and a water puddle below the filter was visible. Photo two shows the product label (lot number and barcode are visible). Photo three shows the same as photo one only from a closer view. The reported condition was verified. The most likely cause of the condition is due to a crack in the housing near the welding zone. A nonconformance has been opened to address this issue. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with u9000 ultrafilter, an external fluid leakage was observed. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameULTRAFILTER HECHINGEN
Type of DeviceSUBSYSTEM, WATER PURIFICATION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
holger - crafoord - strasse 26
hechingen D-723 79
GM D-72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15224235
MDR Text Key298787585
Report Number9611369-2022-00148
Device Sequence Number1
Product Code FIP
UDI-Device Identifier07332414114046
UDI-Public(01)07332414114046
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number114604
Device Lot Number1-1809-H-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/21/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 08/15/2022 Patient Sequence Number: 1
Treatment
AK 98
-
-