Catalog Number 1800300-33 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/25/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that when opening the sterile packaging of the 3.0x33mm xience skypoint stent delivery system the hub was noted to be separated.The device was not used and another xience skypoint was used to successfully complete the procedure.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
A visual inspection was performed on the returned device.The reported material separation was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation was unable to determine a conclusive cause for the reported material separation and noted packaging problem.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
|