Model Number BC606SU
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with bc606su - susi metzenbaum scissors cvd.180mm.According to the complaint description, the "swab forcep" broke off.This occurred during placement of a thoracic drainage tube using the susi-thorax drainage set.An additional medical intervention was required.The piece was recovered.The adverse event / malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: due to the fact that no lot number was provided, a review of the device history records for the complained device is not possible.The review of risk assessment revealed that the overall risk level (severity 2(5) x probability of occurrence 2(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigation results, a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUSI METZENBAUM SCISSORS CVD.180MM
Type of Device
SUSI SINGLE USE SURGICAL INSTR
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key15243053
MDR Text Key298077341
Report Number9610612-2022-00243
Device Sequence Number1
Product Code LRW
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBC606SU
Device Catalogue NumberBC606SU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-