• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR Back to Search Results
Model Number G407209
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2022
Event Type  malfunction  
Event Description
During the procedure.The luer lock created an insufficient seal as air was aspirated into the syringe when attempting to draw back blood.Another needle was use to complete the procedure with no consequences to the patient.
 
Manufacturer Narrative
One brk transseptal needle was received for evaluation.No functional anomalies were noted when the needle was flushed, and an air aspiration test was performed using a slip tip syringe.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the received condition of the device and the information provided, the cause of the reported air remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15254212
MDR Text Key305050473
Report Number3008452825-2022-00561
Device Sequence Number1
Product Code DRC
UDI-Device Identifier05414734205184
UDI-Public05414734205184
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG407209
Device Catalogue NumberG407209
Device Lot Number8496999
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
Patient Weight131 KG
-
-