Catalog Number 682245 |
Device Problem
Break (1069)
|
Patient Problem
Swelling/ Edema (4577)
|
Event Date 07/27/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that 2 bd¿ arterial cannulas' tips broke off during use.The following information was provided by the initial reporter: "complaint came from hospital's medical intensive care unit department.During insertion, catheter tip was breaking.Patient began swelling.At that time, they have tried two cannula and both have the same problem.".
|
|
Manufacturer Narrative
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Manufacturer Narrative
|
H.6.Investigation summary: three photos were received by our quality team for evaluation.From the photos, needle through catheter was observed.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.In the assembly needle line, there is an online lie distance inspection.If the lie distance is out of specification, the part will be automatically rejected by the inspection system.As no sample was received for this complaint, a root cause cannot be determined.H3 other text : see h.10.
|
|
Event Description
|
It was reported that 2 bd¿ arterial cannulas' tips broke off during use.The following information was provided by the initial reporter: "complaint came from hospital micu department.During inserting, catheter tip was breaking.Patient got swealling.At that time they have tried two cannula and both have the same problem.".
|
|