• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD¿ ARTERIAL CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD¿ ARTERIAL CANNULA Back to Search Results
Catalog Number 682245
Device Problem Break (1069)
Patient Problem Swelling/ Edema (4577)
Event Date 07/27/2022
Event Type  malfunction  
Event Description
It was reported that 2 bd¿ arterial cannulas' tips broke off during use.The following information was provided by the initial reporter: "complaint came from hospital's medical intensive care unit department.During insertion, catheter tip was breaking.Patient began swelling.At that time, they have tried two cannula and both have the same problem.".
 
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
H.6.Investigation summary: three photos were received by our quality team for evaluation.From the photos, needle through catheter was observed.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.In the assembly needle line, there is an online lie distance inspection.If the lie distance is out of specification, the part will be automatically rejected by the inspection system.As no sample was received for this complaint, a root cause cannot be determined.H3 other text : see h.10.
 
Event Description
It was reported that 2 bd¿ arterial cannulas' tips broke off during use.The following information was provided by the initial reporter: "complaint came from hospital micu department.During inserting, catheter tip was breaking.Patient got swealling.At that time they have tried two cannula and both have the same problem.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ ARTERIAL CANNULA
Type of Device
ARTERIAL CANNULA
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15254574
MDR Text Key305374260
Report Number8041187-2022-00476
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number682245
Device Lot Number2020987
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-