It was reported that a leakage at the gas outlet connector of the quadrox-i small adult/adult was noticed during patient treatment.The quadrox-i was exchanged during patient treatment.No harm to any person has been reported.The affected product was requested for further investigation but it was not possible to export the product out from china due to custom regulations.Thus a technical investigation could not be performed at the getinge laboratory.Therefore, no probable root cause could be determined.In addition, pictures were provided by the customer on which the reported failure "leakage at the gas outlet connector" is clearly visible.The device history review records were reviewed on 2022-08-15 for the reported failure.Following tests are performed according to the bop as a 100 % inspection: pressure test heat exchanger, leak test water/gas side, pressure test blood side, final functional test.According to the final test results, all 01970110 #squadrox-i ohne filtereinheit with packaging lot#: 3000219279 passed the tests as per specifications.Production related influences are unlikely.Based on the provided pictures by the customer the reported failure "leakage at the gas outlet connector" could be confirmed.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
|