Model Number V173 |
Device Problems
Premature Discharge of Battery (1057); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/29/2022 |
Event Type
Injury
|
Event Description
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) had entered safety mode.This crt-p remains in service.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) had entered safety mode.This crt-p remains in service.No adverse patient effects were reported.Additional information was received and this device has been explanted due to premature battery depletion, and successfully replaced.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption related to telemetry attempts and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) had entered safety mode.This crt-p remains in service.No adverse patient effects were reported.Additional information was received and this device has been explanted due to premature battery depletion, and successfully replaced.No additional adverse patient effects were reported.
|
|
Search Alerts/Recalls
|