Model Number MAJ-1080
Device Problems Leak/Splash (1354); Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/25/2022
Event Type  malfunction  
Event Description
Customer reported leaking on the device and stated machine was inspected and realized the o-ring was missing.The issue found during an unknown event.There was no patient harm, no user injury reported due to the event.
 
Manufacturer Narrative
The subject device was not returned for evaluation.The lot number was not provided and is unknown.Technical assistance center (tac) support engineer communication with the customer noted that tac provided the part number for o-ring and customer will obtain a replacement.No troubleshooting performed as customer will purchase a replacement instead.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see updates to e2, g2, h6 and h10.The device history record was unable to be reviewed for this device since the lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it¿s likely the gas leak was due to the o-ring being disconnected from the hose.The root cause of the disconnected o-ring was unable to be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYLINDER HOSE
Type of Device
CYLINDER HOSE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15272354
MDR Text Key305009194
Report Number3002808148-2022-01270
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170140310
UDI-Public04953170140310
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K014166
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-1080
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-