OGDEN MANUFACTURING PLANT OPTIFLUX 160NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number OPTIFLUX 160NRE DIALYZER FINISHED ASSY
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/15/2022
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A user facility registered nurse reported that a dialyzer blood leak occurred a few minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak.The machine, a fresenius 2008t machine, alarmed appropriately with a ¿blood leak¿ alarm.Blood leak test strips were used and tested negative for the presence of blood, however the leak was visually observed on the arterial end of the dialyzer.There was no defect or damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 200 ml.There was no patient injury or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be available to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
Plant investigation: two complaint companion devices were returned to the manufacturer for physical evaluation.During the gross visual examination of the samples, no defects or irregularities were visually observed.The returned samples were subjected to a laboratory bubble point test.No leaks were detected.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.
 
Event Description
A user facility registered nurse reported that a dialyzer blood leak occurred a few minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak.The machine, a fresenius 2008t machine, alarmed appropriately with a ¿blood leak¿ alarm.Blood leak test strips were used and tested negative for the presence of blood, however the leak was visually observed on the arterial end of the dialyzer.There was no defect or damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 200 ml.There was no patient injury or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be available to be returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX 160NRE DIALYZER FINISHED ASSY.
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key15278077
MDR Text Key305417122
Report Number0001713747-2022-00321
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100149
UDI-Public00840861100149
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162488
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPTIFLUX 160NRE DIALYZER FINISHED ASSY
Device Catalogue Number0500316E
Device Lot Number22HU06008
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received11/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FRESENIUS 2008T MACHINE; FRESENIUS 2008T MACHINE; FRESENIUS BLOODLINES; FRESENIUS BLOODLINES
Patient Age86 YR
Patient SexFemale
Patient Weight71 KG
-
-