Corrected fields: b5, d5, e1(event site name: (b)(6) hospital , event site address: (b)(6), event site city: (b)(6)).Additional information: e1(event site postal code: (b)(6), event site email: (b)(6)).It was reported that during a routine check the cs300 intra-aortic balloon pump (iabp) was not switching on the first time and could take 4 attempts to get the unit to switch on.A getinge field service engineer (fse) was dispatched to investigate the issue.The fse advised the customer that a delay is required between multiple on/ off actions.The fse then noticed that the on-off switch was not switching clearly.So, in order to fix the issue, the fse replaced the cable assembly.The unit passed all functional and calibration tests.The unit was then put back into clinical use.
|