Model Number 3660 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 06/01/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event is estimated.
|
|
Event Description
|
It was reported that the patient underwent an unrelated surgical procedure.It's unknown if patient set the ipg into surgery mode as recommended.Thereafter, the ipg failed to establish communication with external device.Recovery efforts were unsuccessful and the ipg was deemed inoperable.Patient is not receiving therapy and may require surgical intervention to address the issue.
|
|
Manufacturer Narrative
|
An inoperable pulse generator was reported to abbott.The system was not set to surgery mode while the patient underwent an unrelated surgery where electro-cautery may have been used.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.The results of the investigation are inconclusive since the device was not returned for analysis.Actions have been taken to prevent reoccurrence.
|
|
Event Description
|
Additional information received identified that the ipg was successfully recovered.Effective therapy restored.
|
|