DAVOL INC., SUB. C.R. BARD, INC. -1213643 XENMATRIX PORCINE SURGICAL MESH Back to Search Results
Catalog Number 1161935
Device Problem Patient-Device Incompatibility (2682)
Patient Problems Wound Dehiscence (1154); Fistula (1862); Hernia (2240)
Event Date 07/31/2015
Event Type  Injury  
Manufacturer Narrative
As reported, post implant of the xenmatrix graft , the patient was diagnosed with a recurrent hernia, wound dehiscence and fistula and underwent wound debridement, fistulotomy & mesh explant surgery. The clinician has assessed the patient¿s postoperative course as possibly related to the study device and not related to the procedure. However; based on the information provided, no conclusion can be made. Hernia recurrence and fistula are known inherent risks of surgery and are listed as possible complications in the instructions-for-use (ifu) supplied with the device. Review of manufacturing records confirms product was manufactured to specification. To date, this is the only reported complaint for this manufacturing lot of (b)(4) units released for distribution in january 2015. This mdr represent the ae of hernia recurrence, fistula, wound dehiscence for the xenmatrix (device #1). An additional mdr has been submitted to represent the ae of hernia recurrence for the xenmatrix (device #2). The information provided by bd represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
Per clinical trial (b)(4): on (b)(6) 2015, the patient underwent primary open midline ventral hernia repair during which xenmatrix surgical graft (device #1) was trimmed and placed using underlay technique in intra-peritoneal region and fixated using sutures. Full skin closure was achieved using maxon sutures with interrupted stitch. The patient was discharged (b)(6) 2015. As reported, the patient was diagnosed with wound dehiscence (b)(6) 2015 and admitted in the hospital. On (b)(6) 2015, the patient underwent surgery for wound dehiscence and given iv antibiotics, debridement, wound care and wound vac was placed. The patient was discharged on (b)(6) 2015. As reported, the patient was diagnosed with enterocutaneous fistula on (b)(6) 2015 and admitted in the hospital. As reported the patient was treated with tpn (total parenteral nutrition). The patient was discharged (b)(6) 2015. As reported, (b)(6) 2015. The patient was diagnosed with midline ventral hernia recurrence and ec fistula. As reported, the patient underwent surgery for hernia recurrence, wound dehiscence & fistula on (b)(6) 2015 during which another xenmatrix surgical graft, (device #2) was implanted and fixated using sutures and the previously implanted xenmatrix mesh (device #1) was removed (100%) and also fistulotomy occurred at the same time. The patient was discharged to home (b)(6) 2015. As reported (b)(6) 2017. The patient was diagnosed with midline small ventral hernia recurrence (device #2). The patient decided not to surgically treat at that time. The patient returned a year later and decided to repair again and underwent ct scan on (b)(6) 2018. Per the study clinician, the reported ae (hernia recurrence, fistula, wound dehiscence) has been assessed as moderate in severity, possibly to the study device and not related to the index procedure and recovered/resolved. The reported ae does meet the definition of serious (sae) and does not meet an unlisted ade (uade).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameXENMATRIX
Type of DevicePORCINE SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key15282770
MDR Text Key298479114
Report Number1213643-2022-00592
Device Sequence Number1
Product Code FTM
UDI-Device Identifier00801741031441
UDI-Public(01)00801741031441
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Study
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Expiration Date12/28/2016
Device Catalogue Number1161935
Device Lot NumberHUYJ0178
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 08/24/2022 Patient Sequence Number: 1
-
-