COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
|
Back to Search Results |
|
Model Number G50333 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
As reported, during a tibial angioplasty procedure, an advance 14 lp low profile balloon catheter leaked upon the first inflation.The balloon leaked and blood was noted in the balloon lumen.Another balloon was used to complete the procedure.There were no adverse effects to the patient.
|
|
Event Description
|
Upon return and initial evaluation of the device on 26aug2022, the device was found to be leaking below the strain relief/hub.Although a leak was not noted in the balloon material, the balloon was not able to be inflated.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional/corrected information: b5, h3: (other): device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed, h6 (annex g).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
No additional information has been received since the last report was submitted.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, during a tibial angioplasty procedure, an advance 14 lp low profile balloon catheter leaked upon the first inflation.The balloon leaked and blood was noted in the balloon lumen.Another balloon was used to complete the procedure.There were no adverse effects to the patient.Upon return and initial evaluation of the device on (b)(6) 2022, the device was found to be leaking below the strain relief/hub.Although a leak was not noted in the balloon material, the balloon was not able to be inflated.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.The balloon did not inflate.A leak was noted under the strain relief.The strain relief was pulled down, and the device leaked from the hub, at the location of the glue mound.No hub damage was noted.A document-based investigation evaluation was performed.There have been no other reported complaints for this lot number.One relevant non-conformance was noted on a sub-assembly lot; however, all affected product was scrapped.Although the non-conformance is relevant to the reported failure mode, all non-conforming product was scrapped, there are 100% inspections to capture this non-conformance, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints have been received from the field.Therefore, it was concluded that there is no evidence that non-conforming product exists in-house or in field.The product ifu states ¿upon removal from package, inspect the product to ensure no damage has occurred.¿ the information provided upon review of the dhr, dmr, ifu and investigation of the returned device suggests that the complaint device was built out of specification; however, cook has concluded that there is no evidence that additional non-conforming product exists in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that a quality control deficiency contributed to the failure mode.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|