During a procedure today with dr.(b)(6) and dr.(b)(6) (right acl repair) the smith and nephew fluid pump malfunctioned.Rn noticed that pump pressure inadvertently increased to 105, upon noticing this surgeon was immediately notified and the pump was lowered to the proper pressure of 50.The surgeon noticed some edema in the patient's right knee.Upon arrival in the pacu the patient's edema was significantly decreased.The piece of equipment was removed from the or and tagged broken.Representative was notified regarding malfunction and will be in to verify issue.The pump was changed out for another, the cartridge was changed, bio-med and the vendor for the pump were called.There was nothing found to be not working appropriately with the pump.The vendors are on site most days to support any potential issues.This issue was discussed in our staff meeting to further educate and to make all staff aware to observe for any further concerns.
|