Device Problem
Device Alarm System (1012)
|
Patient Problems
Low Oxygen Saturation (2477); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/04/2022 |
Event Type
Injury
|
Event Description
|
It was reported that the device is showing no disposable alarm.No patient injury was reported.
|
|
Manufacturer Narrative
|
No information has been provided to date.No lot number was provided; therefore, device history record review could not be completed.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed as no product sample was received for evaluation.If the product is returned, the manufacturer will reopen this complaint for further investigation.
|
|
Manufacturer Narrative
|
Corrected data: b5 describe event or problem, h1 type of reportable event, and h6 patient code.
|
|
Event Description
|
Troubleshooting of the pump did not resolve the no disposable alarm.A new cassette was used with the pump and the infusion continued without issue.The oxygen saturation of the patient at the time of this event was 85 percent which is below the normal level.The patient was given oxygen via nasal cannula and the oxygen saturation returned to a normal level.While on the phone, the patient began resting more comfortably and the oxygen saturation level increased to 91 percent.
|
|