• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. RESERVOIR 3ML MMT-332A; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. RESERVOIR 3ML MMT-332A; PUMP, INFUSION Back to Search Results
Model Number MMT-332A
Device Problems Insufficient Flow or Under Infusion (2182); Obstruction of Flow (2423)
Patient Problems Hyperglycemia (1905); Polydipsia (2604)
Event Date 08/17/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the insulin pump was under delivering the insulin.Customer stated the reason the insulin pump was under delivering the insulin was the customer was insulin dependent and thought it was site.The customer's blood glucose was 33 mmol/l and the current blood glucose was 24.6 mmo/l.Customer treated high blood glucose with insulin pump.The insulin pump alarmed insulin flow block alarm.Customer stated that there was no leakage but they could smell the insulin.No further patient complications were reported.Troubleshooting was performed.The device will not be returned for the analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESERVOIR 3ML MMT-332A
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key15301608
MDR Text Key298688181
Report Number2032227-2022-316087
Device Sequence Number1
Product Code FRN
UDI-Device Identifier000020763000273358
UDI-Public(01)000020763000273358(17)231008(10)HG4R9JB
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/08/2023
Device Model NumberMMT-332A
Device Catalogue NumberMMT-332A
Device Lot NumberHG4R9JB
Was Device Available for Evaluation? No
Date Manufacturer Received08/17/2022
Date Device Manufactured10/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-