The reported observation of ¿ipg communication¿ was confirmed.The analysis results concluded the inability to establish communication between the programmer and the implantable pulse generator (ipg) was due to the device entering the service application state.The root cause of the device entering the service application state was due to the patient¿s procedure.Per the clinicians manual arten600266417 warnings - electrosurgery.To avoid harming the patient or damaging the neurostimulation system, do not use monopolar electrosurgery devices on patients with implanted neurostimulation systems.Before using an electrosurgery device, place the device in surgery mode using the patient controller app or clinician programmer app.Confirm the neurostimulation system is functioning correctly after the procedure.Under warnings - there is sufficient documentation concerning the use of use short-wave diathermy, microwave diathermy, or therapeutic ultrasound diathermy (all now referred to as diathermy) on patients implanted with a neurostimulation system, and medical procedures (such as therapeutic radiation and electromagnetic lithotripsy).The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on (b)(6)2017.
|