Model Number 407439
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
One 8f swartz braided introducer sheath and dilator were received for evaluation. The sheath and dilator had been perforated proximal to the distal tip. A needle/stylet assembly from current inventory was inserted and advanced through the dilator/sheath assembly; however, the needle could not advance past the location of the perforation. The device history record was reviewed to ensure that each manufacturing and inspection operation was performed. The cause of the device perforation remains unknown.
 
Event Description
This report is to advise of a punctured dilator and sheath noted during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER SL 1 CURVE, 63 CM LENGTH, 8F
Type of DeviceINTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15303411
MDR Text Key299405797
Report Number3005334138-2022-00491
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734205634
UDI-Public05414734205634
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K052644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation
Type of Report Initial
Report Date 08/26/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number407439
Device Catalogue Number407439
Device Lot Number8103562
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-