• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; STYLET, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; STYLET, CATHETER Back to Search Results
Model Number 518-062
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 07/29/2022
Event Type  Injury  
Manufacturer Narrative
A portion of the device was discarded and a portion remained in the patient, thus no investigation could be completed.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event Description
A lead extraction procedure commenced to remove a right ventricular (rv) lead due to cied system/pocket infection.A spectranetics lead locking device (lld) was inserted into the lead to provide traction.Using spectranetics tools including a glidelight laser sheath and a tightrail rotating dilator sheath, the physician was unable to remove the rv lead.He attempted to unlock the lld from the lead without success.The rv lead and the lld were cut and capped, and abandoned in the patient.This report captures the lld present within the rv lead which was cut and capped, and abandoned in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
STYLET, CATHETER
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key15303845
MDR Text Key298717220
Report Number1721279-2022-00161
Device Sequence Number1
Product Code DRB
UDI-Device Identifier20813132023076
UDI-Public(01)20813132023076(17)240404(10)FLP22C31A
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberFLP22C31A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MEDTRONIC 6935 RV ICD LEAD; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; SPECTRANETICS GLIDELIGHT LASER SHEATH; SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH
Patient Outcome(s) Other;
Patient Age70 YR
Patient SexMale
Patient Weight120 KG
-
-