Model Number FG15160-0615-1S |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Hemorrhage/Bleeding (1888)
|
Event Date 07/21/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that a patient treated with a ped3 pipeline and phenom 27 catheter had active bleeding on the right femoral artery that required prolonging of hospitalization.Bleeding on the right common femoral artery distal to the origin of the circumflex artery.The event resolved (b)(6) 2022.The event had a causal relationship to the catheter and the study procedure and a probable relationship to the antiplatelet medication.The ped3 was successfully implanted and wall apposition was achieved.The patient was being treated for an aneurysm. side (hemisphere): left.Location (vessel): internal carotid artery.Location of aneurysm in vessel: sidewall.Aneurysm morphology: saccular.Maximum aneurysm diameter: 4.7mm.Aneurysm dome height: 3.7mm.Aneurysm dome width: 4.7mm.Aneurysm neck size: 3.4mm.Parent artery diameter distal to aneurysm: 3.9mm.Parent artery diameter proximal to aneurysm: 3.2mm.Significant stasis at the end of the procedure.Complete neck coverage at the end of the procedure.Raymond and roy occlusion class 3 at the end of the procedure.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Additional information received reported that per source, 21jul2022 abd/pelvic scan, hematoma of rt internal oblique muscle measuring 29x85x76 mm, w/active leak fed by arterial branch involving thehomolateral external iliac artery.Active bleeding on the right fem oral artery (vascular access site) was reported.
|
|