ST PAUL BCI WW1030 PULSE OXIMETER Back to Search Results
Model Number WW1030P1EN
Device Problems Defective Alarm (1014); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review was not performed because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.The device was returned for evaluation.Visual and functional tests were performed.The device was received in good physical condition but the red light-emitting diode (led) was damaged from impact.The reported problem was duplicated.The monitor was powered up and the red led did not light up.The main board was replaced.The root cause of the reported issue was found to be impact to the monitor.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that the light on top of the device will not work when alarming.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI WW1030 PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
3223 32nd ave s
minneapolis, MN 55442
MDR Report Key15321541
MDR Text Key305086673
Report Number3012307300-2022-16792
Device Sequence Number1
Product Code DQA
UDI-Device Identifier30843418001211
UDI-Public30843418001211
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083557
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWW1030P1EN
Device Catalogue NumberWW1030P1EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2020
Was the Report Sent to FDA? No
Date Manufacturer Received05/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-