Lot Number 191226 OR 200211
Device Problem Disconnection (1171)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2020
Event Type  malfunction  
Manufacturer Narrative
The sample was returned from the customer for evaluation.When observing the male connector of the filter, it was confirmed that the retention edge which is designed to hold rotating collar connected with filter was damaged.As a result of the investigation, it was confirmed that the retortion edge was damaged and flattened, and it is possible that this event was caused.The timing of the collapse of the retortion edge and the cause of the reported event could not be identified.The occurrence of this event will be notified to our manufacturing department and this event will be recorded in the database to monitor the future occurrence.Device history record review of manufacturing records, relevant to the lot reported, found no discrepancies or anomalies.(b)(4).
 
Event Description
It was reported that when the epifuse connector was connected with the filter, the connection part broke.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX EPIDURAL CONTINUOUS TRAYS
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer Contact
jim vegel
6000 nathan lane north
3-10-15, hakata station east,
minneapolis, MN 55442
MDR Report Key15322063
MDR Text Key305377084
Report Number3012307300-2022-16812
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K781052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Lot Number191226 OR 200211
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/13/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-