A device history record (dhr) review was conducted which indicated all inspection were completed and no issues were noted during manufacture.Upon observing the returned product, it was confirmed that the stylet was bent at the base of the cap.The reported event was confirmed; however, the root cause could not be established.The manufacturer will continue to monitor the future occurrence status.No product information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
|