• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL INTRODUCER, CATHETER Back to Search Results
Model Number D138501
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Manufacturer Narrative
The exact date of event is unknown and therefore, this field was populated with (b)(6)2022. The biosense webster, inc. Product analysis lab received the device for evaluation. The analysis has begun but is not completed at this time. When the investigational analysis has been completed, a supplemental 3500a report will be submitted. This report is being submitted pursuant to the provisions of 21 cfr, part 4. This report may be based on information which has not been investigated or verified prior to the required reporting date. This report does not reflect a conclusion by biosense webster, inc. , or its employees that the report constitutes an admission that the product, biosense webster, inc. , or its employees caused or contributed to the potential event described in this report. If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate. Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent a cardiac ablation procedure with a carto vizigo¿ 8. 5f bi-directional guiding sheath ¿ small and a hemostatic valve separation issue occurred. It was reported that there was a hemostatic valve defect. Blood came out of the valve. Back bleeding was noticed, valve probably dislodged. The sheath was being used on the patient. Air did not enter the patient¿s body. This issue did not require percutaneous or surgical removal. The patient¿s hemodynamics were not compromised due to bleeding. No medical intervention was required to stop the bleeding. There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL
Type of DeviceINTRODUCER, CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (IRVINE)
33 technology drive
irvine CA 92618
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-798
MDR Report Key15325184
MDR Text Key304299742
Report Number2029046-2022-02045
Device Sequence Number1
Product Code DYB
UDI-Device Identifier10846835016253
UDI-Public10846835016253
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K170997
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/19/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberD138501
Device Catalogue NumberD138501
Device Lot Number00001919
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received09/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-