• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BCI CAPNOCHECK PLUS; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BCI CAPNOCHECK PLUS; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number 9004050
Device Problem High Readings (2459)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review was not performed because the results of the complaint investigation do not indicate a problem with the manufacturing of the device or prior repair of the device.A product sample was received for evaluation.Visual inspection and functional testing were performed.No physical damage was observed.The reported device was put on the carbon dioxide (co2) simulator for testing.The device was getting high end-tidal co2 (etco2) reading of 79mmhg.However, the range for testing was set between 70 to 76 mmhg.The technician was able to duplicate the reported issue of high readings.The technician performed the high/low calibration to resolve the reported issue.After calibration, the technician was able to get the etco2 reading of 73 mmhg, which is within specifications.The root cause of the reported issue was determined to be related to the customer/ user interfacing with the product in a manner inconsistent with the instruction for use (ifu).The ifu in the operator manual recommends the user perform high/low calibration every 6 months to ensure that the etco2 and o2 measurements are accurate.No product information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
 
Event Description
It was reported that there was an high readings issue with the reported device.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI CAPNOCHECK PLUS
Type of Device
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
1195 w fremont ave
minneapolis, MN 55442
MDR Report Key15326684
MDR Text Key305346391
Report Number3012307300-2022-16925
Device Sequence Number1
Product Code CCK
UDI-Device Identifier10610586036866
UDI-Public10610586036866
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K970209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number9004050
Device Catalogue Number9004050
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2020
Was the Report Sent to FDA? No
Date Manufacturer Received05/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-