It was reported that the real intelligence tracking camera lenses crack and the real intelligence cori do not function, boot software; it appears that the screen remain black and console remain flickering with orange light instead of normal console lighting.As this was noticed during set up, patient was not impacted.
|
Section h3, h6: the real intelligence cori, part number rob10024, serial number (b)(6) intended for treatment was not returned for evaluation.An image was provided.A relationship between the reported event and the device was established.The reported problem was confirmed with a visual inspection.The image provided confirmed the camera is cracked and the screen remains black.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release a serial number, lot number, part revision or software version is required to link the device to a dhr or nc investigation.The failure mode and associated risk have been anticipated within the risk file.The risk level is still adequate.A historical review concluded that no prior escalation actions are applicable to the scope of the reported complaint.The most likely cause of this event is associated with a software bug.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|