Model Number ROB10024
Device Problem Computer Operating System Problem (2898)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/12/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that the real intelligence tracking camera lenses crack and the real intelligence cori do not function, boot software; it appears that the screen remain black and console remain flickering with orange light instead of normal console lighting.As this was noticed during set up, patient was not impacted.
 
Manufacturer Narrative
Section h3, h6: the real intelligence cori, part number rob10024, serial number (b)(6) intended for treatment was not returned for evaluation.An image was provided.A relationship between the reported event and the device was established.The reported problem was confirmed with a visual inspection.The image provided confirmed the camera is cracked and the screen remains black.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release a serial number, lot number, part revision or software version is required to link the device to a dhr or nc investigation.The failure mode and associated risk have been anticipated within the risk file.The risk level is still adequate.A historical review concluded that no prior escalation actions are applicable to the scope of the reported complaint.The most likely cause of this event is associated with a software bug.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15341972
MDR Text Key304791100
Report Number3010266064-2022-00590
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757420
UDI-Public00885556757420
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10024
Device Catalogue NumberROB10024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-