Model Number 406849
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2022
Event Type  malfunction  
Event Description
During the atrial fibrillation procedure, after being inserted, air was aspirated from the sheath when drawing back the syringe.It was noted that the sheath continued to produce air so the sheath was exchanged.The procedure was completed with no adverse patient health consequences.
 
Manufacturer Narrative
Additional information: d9, g3, h2, h3.One 8.5f fast-cath introducer sheath was received for evaluation.The dilator was also received.The sheath passed pressure and aspiration leak testing with no anomalies observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported leak remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15348364
MDR Text Key299406030
Report Number3005334138-2022-00499
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734203388
UDI-Public05414734203388
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number406849
Device Catalogue Number406849
Device Lot Number8524452
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-