|
Model Number 29420 |
Device Problem
Entrapment of Device (1212)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr: a visual analysis of the returned device revealed a guidewire was stuck in the device.The renegade device showed multiple bends and kinks on the shaft, and was severely stretched and separated.The location of the separation was 94cm from the hub.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2022.It was reported that the microcatheter became stuck inside the catheter during advancing.A 135/20 renegade hi-flo was selected for use in the liver.Prior to the procedure, it was noted that the microcatheter became stuck inside the catheter during advancing.The procedure was completed with another of same device.No complications were reported and patient was stable post procedure.However, device analysis revealed that there was separation located at 94cm from the hub.
|
|
Search Alerts/Recalls
|
|
|