|
Model Number MMT-105ELGYNA |
Device Problems
Break (1069); Patient Device Interaction Problem (4001)
|
Patient Problem
Hyperglycemia (1905)
|
Event Date 08/18/2022 |
Event Type
malfunction
|
Event Description
|
Information received by medtronic indicated that the insulin pen was not working.Customer reported that insulin pen lead screw was not intended.The screw retracted when dose dialing.No harm requiring medical intervention was reported.The customer will discontinue to use the device.
|
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Per visual inspection: no physical damage to cartridge holder, car or inpen was noted.Customer reports: the pushing mechanism its not pushing and it retracts.Sugars are also high.Unit paired successfully to commercial app.My inpen menu displayed: the following test values were dialed and dosed: 10 test values of 4.0u dialed and dosed.All values displayed accurately in the logbook.All doses were written to the app logbook within 1 minute.Inpen received with 1/4 of the travel.The screw was not bent, advanced when dosage knob was pressed dialing a dosage and retracted appropriately.No resistance was observed when dosing without a cartridge installed.The screw advanced every time 30.0u was dialed and dosed until the screw reached max extension.In conclusion: inpen received working as design.No malfunctions noted during testing that could affect insulin delivery.The customer complaint of difficult button not pushing insulin or leadscrew anomaly could not be confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|
|
|