MAUDE Adverse Event Report: COVIDIEN LP - SUPERDIMENSION INC ILLUMISITE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP - SUPERDIMENSION INC ILLUMISITE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number ILS-1800-KT
Device Problems Device Displays Incorrect Message (2591); Application Program Problem (2880)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 07/11/2022
Event Type  Injury  
Manufacturer Narrative
Out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.Concomitant medical products: ils-1800-kt ils-1800-kt procedure kit illumisite 180 lot # 519926.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a procedure, the connection with locatable guide (lg) kept flickering, disconnecting and reconnecting rapidly.Another catheter was used but have the same issue.The devices were indicated as outside of the sensor sensing volume. the physician was not able to complete the electromagnetic navigational bronchoscopy (enb) portion of the case, even by other means.The patient was under general anesthesia.
 
Manufacturer Narrative
Additional information: a5a, a5b, d9, g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had excessive adhesive.It was reported that the patient procedure was cancelled or aborted due to a system issue and the system displays error message indicating lg or pst is outside the magnetic volume.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H6 patient codes: e2402 (cancelled case) additional info: g3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that there was an excessive amount of adhesive on the tip of the device.Functionally, the extended working channel (ewc) length was measured using a calibrated ruler and was found to be within specification.The ewc continuity was checked using a breakout box and calibrated digital multimeter, and the results were acceptable.The ewc sensor wire was checked for shorts and there were no shorts found.The locatable guide (lg) length was measured using a calibrated ruler and was found to be within specification.The lg continuity was checked with a breakout board and the results were acceptable.It was reported that the patient procedure was cancelled or aborted due to a system issue and the system displays error message indicating lg or pst was outside the magnetic volume.The reported issues of outside the magnetic volume could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: the unit had excessive adhesive.The most likely cause for the excessive adhesive was determined to be manufacturing related.Internal process improvements have been initiated to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILLUMISITE
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer (Section G)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key15397276
MDR Text Key299663693
Report Number3004962788-2022-00074
Device Sequence Number1
Product Code JAK
UDI-Device Identifier10884521723818
UDI-Public10884521723818
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K191394
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 12/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberILS-1800-KT
Device Catalogue NumberILS-1800-KT
Device Lot Number519926
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
SEE H10.
Patient Outcome(s) Other;
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-