Catalog Number 2C8750 |
Device Problems
Clumping in Device or Device Ingredient (1095); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/09/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a clot made it through the filter of a clearlink system y-type blood/solution set.A tear was discovered on the filter, approximately 1 inch long.This issue was identified during patient use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed observe a tear in the filter.The reported condition was verified.The cause of the condition was due to the supplier's manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|