The log file evaluation revealed that the device forced a shut-down of automatic ventilation during the procedure in question after having detected a deviation of the ventilator motor's rotation speed.To protect the patient from potentially hazardous output, motor position, rotation speed and acceleration must be monitored continuously.If the measured values have a significant deviation from the ones calculated by the software, the system is designed to shut-down automatic ventilation for safety reasons.The user will be alerted by means of a corresponding alarm; manual ventilation including monitoring remain available.Based on the entries found the cause can be traced back to the ventilator motor or the position detection system.Nevertheless, during on-site evaluation a faulty vacuum pump was determined as root cause.After replacement of the pump the problem was initially solved until it occurred again in the repair workshop two months later.The motor, pneumatic assembly and main board were then replaced.All replaced components were requested and sent, but only the motor was received for investigation.The motor was subject to an in-depth evaluation in the manufacturer's lab but the reported ventilator failure could not be reproduced.Therefore, the exact root cause could not be determined.Finally, it can be concluded that the device has reacted to the situation as specified.
|