The icd and fragments of the leads were returned for analysis.Upon receipt, the leads were still connected to the icd header.Incoming visual inspection showed no anomalies.In a next step, the leads were disconnected from the icd and the devices were analyzed.Incoming icd interrogation revealed the battery status bos.No charging cycles were documented.The memory content of the icd was inspected, showing a normal device function while implanted and in service.The analysis revealed that a magnet application was detected by the device on (b)(6) 2022.As a result, a message box was shown on the programmer screen at first interrogation on (b)(6) 2022 at 10:00h, informing about the magnet application and asking whether the anti-tachycardia detection should be disabled.This message box was closed by pressing the yes button at 10:32h, which led to the clinical observation.There was no indication of a device malfunction.The icd was subjected to an electrical analysis.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied, and the device delivered a defibrillation shock as specified, documenting a correct sensing and shock delivery.In particular, the specified energy level was reached.In conclusion, the icd proved to be fully functional.The analysis of the memory content showed a normal device behavior while the device was implanted and in service.There was no indication of an icd malfunction.The received leads were found cut next to the connector pin.The proximal fragment of the leads were received for analysis.It is reasonable to assume that the leads were cut during the explantation procedure.The insulation of the leads was apart from the present cut, free of breaches.Further inspection of the leads did not show any deviations that might have contributed to the clinical observations.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|