Catalog Number EMC7109 |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/17/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter first name: initial reporter last name: initial reporter facility name: initial reporter address: initial reporter city: should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the connector of a two (2) female luer lock adapters with control-a-flo were cracked.The issue was identified during patient infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The lot was manufactured in february 2021.The actual devices were not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph which showed a crack at the female connector.The reported condition was verified.The cause of the condition was due to a supplier related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|