Catalog Number EMC7109
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter first name: initial reporter last name: initial reporter facility name: initial reporter address: initial reporter city: should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the connector of a two (2) female luer lock adapters with control-a-flo were cracked.The issue was identified during patient infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The lot was manufactured in february 2021.The actual devices were not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph which showed a crack at the female connector.The reported condition was verified.The cause of the condition was due to a supplier related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I.V. ADMINISTRATION SETS WITH CONTROL-A-FLO REGULATOR
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - TUNISIA
route de chebbaou
2021oued e
tunis
TS  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15427899
MDR Text Key301244160
Report Number1416980-2022-04843
Device Sequence Number1
Product Code FPA
UDI-Device Identifier05413760192321
UDI-Public(01)05413760192321
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMC7109
Device Lot Number21B03T430
Was Device Available for Evaluation? No
Date Manufacturer Received10/14/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-