• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL EXPRESSEW III AUTOCAPTURE + SUTURE PASSER; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL EXPRESSEW III AUTOCAPTURE + SUTURE PASSER; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Model Number 288233
Device Problems Device-Device Incompatibility (2919); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
During surgery the auto capture device it was reported by a healthcare professional in india that during a rotator cuff repair procedure on an unknown date, it was observed that the expressew iii autocapture + suture passer device was getting locked & not retrieving/opening.During in-house engineering evaluation, it was determined that the upper jaw on the device was slightly loose when the jaws were closed; therefore, it showed that the jaws did not close normally contributing to the holding and close issue.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The device manufacture date is currently unavailable.Investigation summary the product was returned to mitek for evaluation.Mitek then conducted visual inspection and functional test of device received by the customer.Upon visual inspection, the device has signs of use;.To test its functionality, it was tested on a sample rubber strip; as a result, the upper jaw was slightly loose when the jaws are closed; therefore, it showed that the jaws did not close normally contributing to the holding and close issue.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.According with the visual inspection and the functional test result, this complaint can be confirmed.The possible root cause for the issue experienced can be attributed when clamping excessive tissue between the jaws combined with repetitive twisting action exerts excess load on the jaws causing it to eventually damage the jaw.Since this is a reusable device, this failure has possibly occurred after using it in this manner for many procedures.As per ifu, it is important to inspect the device prior to use to ensure proper mechanical function.Visually inspect the instrument and check for damage and wear.Also, jaws and teeth should align properly.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESSEW III AUTOCAPTURE + SUTURE PASSER
Type of Device
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15427953
MDR Text Key299989242
Report Number1221934-2022-02846
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10886705027101
UDI-Public10886705027101
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number288233
Device Catalogue Number288233
Device Lot Number61822-210612-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/01/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-