The device manufacturer date is not known.This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record attached, the reported failure "damage on the distal end of the scope, causing fogging" was confirmed.According to henke: scope evaluated as a level 3.Dents on needle, bent needle, distal tip/fiber damage, shaver damage to distal tip, chipped lens in system, scratches on needle, residue on external optics.Probably root cause for this failure is: the complaint for ¿damage on the distal end of scope causing fogging¿ has been confirmed and is due to customer use and handling.The reported failure mode will be monitored for future reoccurrence.
|