• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number C37101307J
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/21/2020
Event Type  malfunction  
Manufacturer Narrative
Udi information is unknown.Operator of device is unknown.No information has been provided to date.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.The circuits did not pass the leak test.Unable to determine the root cause of the reported issue.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#(b)(4).
 
Event Description
It was reported that during a pre-use air leak check, the customer found leakage of air from the product.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
4-1-5, izumi-chuo, izumi-ku,
minneapolis, MN 55442
MDR Report Key15433421
MDR Text Key306071287
Report Number3012307300-2022-19336
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K033080
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date01/28/2023
Device Catalogue NumberC37101307J
Device Lot Number3930796
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2020
Was the Report Sent to FDA? No
Date Manufacturer Received07/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-