BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® K3E 5.4MG PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 368856
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to stopper pop off (decapping) as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode stopper pop off (decapping).Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that while using bd vacutainer® k3e 5.4mg plus blood collection tubes the customer observed that the cap of the tube jumps.This event occurred 10 times.The following information was provided by the initial reporter: "the cap of the tube jumps[.] it has happened on several occasions that after sampling the cap of the tube "jumps".".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K3E 5.4MG PLUS BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15440415
MDR Text Key302763086
Report Number9617032-2022-00902
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Catalogue Number368856
Device Lot Number2117503
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-