MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX TUBES PDT ULTRAPERC; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX TUBES PDT ULTRAPERC; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 100/596/090
Device Problem Labelling, Instructions for Use or Training Problem (1318)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2021
Event Type  malfunction  
Manufacturer Narrative
No product sample has been provided.The customer provided a photo for evaluation.The photo showed an original product with green (7mm) size block in the upper left corner.The condition observed on the photo does not meet the requirements of labeling artwork for product.Device history record review for the affected lot found that product includes 9 mm trach tube, which corresponds with blue (9 mm) size block on the artwork.Based on the provided evidences the reported by the customer allegation was confirmed.(udi) and pma/510k are unknown.No information is provided to date.Operator of device is unknown.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that the device size is labeled incorrectly.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES PDT ULTRAPERC
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15445134
MDR Text Key306072215
Report Number3012307300-2022-19586
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date12/01/2021
Device Catalogue Number100/596/090
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1993-2021
Patient Sequence Number1
-
-