BOSTON SCIENTIFIC NEUROMODULATION; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
|
Back to Search Results |
|
Model Number NM-3138-55 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Scar Tissue (2060); Discomfort (2330); Swelling/ Edema (4577)
|
Event Date 09/07/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional suspect medical device component involved in the event: product family: dbs-extension, upn: (b)(4), model: nm-3138-55, serial: (b)(4), batch: 20979566.
|
|
Event Description
|
It was reported that the patient experienced discomfort at the retro-auricular area, where the patient repetitively touched and moved the connection of the leads and extensions.The physician observed swelling and scarring in the area.The patient underwent a lead extension revision procedure.During the procedure, the physician noticed the extensions were knotted so the extensions were replaced.The patient was doing well post-operatively.The explanted lead extensions were kept by the facility.
|
|