• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367324
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2022
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this.There were multiple device types reported to be involved.The information for the additional device types are as follows.Medical device type: jka / fpa.Common device name: blood specimen collection device; intravascular administration set a device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported when using the bd vacutainer® push button blood collection set, the device experienced cut tubing.This event occurred once.The following information was provided by the initial reporter.The customer stated: the customer reported about cut tubing and preactivation.
 
Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for evaluation: yes d9: returned to manufacturer on: 2022-10-20 h.6.Investigation summary: bd received 2 samples and 6 photos for investigation.The photos were reviewed and the customer¿s indicated failure modes of cut tubing and preactivation was observed.Additionally, the customer samples were evaluated by visual examination and the indicated failure modes of cut tubing and preactivation with the incident lot was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure modes cut tubing and preactivation.Bd determined that the root cause of the indicated failure mode of cut tubing was attributed to manufacturing.Bd was not able to identify a root cause for the indicated failure mode of preactivation.H3 other text : see h.10.
 
Event Description
It was reported when using the bd vacutainer® push button blood collection set, the device experienced cut tubing.This event occurred once.The following information was provided by the initial reporter.The customer stated: the customer reported about cut tubing and preactivation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15488874
MDR Text Key305515181
Report Number1710034-2022-00545
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903673242
UDI-Public50382903673242
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K030573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/31/2023
Device Model Number367324
Device Catalogue Number367324
Device Lot Number1246701
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-