MAUDE Adverse Event Report: BIOSENSE WEBSTER INC CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL; INTRODUCER, CATHETER Back to Search Results
Model Number D138501
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(6).This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.(b)(4).
 
Event Description
It was reported that a patient underwent a cardiac ablation procedure with a carto vizigo¿ 8.5f bi-directional guiding sheath ¿ small and a hemostatic valve separation issue occurred.It was reported that there was a hemostatic valve defect.The hemostatic valve fell into the hub.The issue was confirmed at insertion into sheath.The issue was resolved by changing the vizigo¿ to another one.The procedure was then successfully completed without patient consequence.
 
Manufacturer Narrative
The device evaluation was completed on 23-nov-2022.It was reported that a patient underwent a cardiac ablation procedure with a carto vizigo¿ 8.5f bi-directional guiding sheath ¿ small and a hemostatic valve separation issue occurred.It was reported that there was a hemostatic valve defect.The hemostatic valve fell into the hub.The issue was confirmed at insertion into sheath.The issue was resolved by changing the vizigo¿ to another one.The procedure was then successfully completed without patient consequence.Device evaluation details: the product was returned to biosense webster inc.(bwi) for evaluation.Visual inspection and microscopic examination of the returned device were performed following bwi procedures.Visual analysis revealed that the hemostatic valve was broken inside the hub component.Microscopic examination of the hemostatic valve surface showed stress marks on the outer diameter.The stress marks suggest that excessive force or manipulation was applied due to an extreme off axis angle of insertion.Valve damage occurs when extreme off axis angles is performed during insertion with the dilator, outside of what is recommended in the odp (optimal device performance guide).A device history record (dhr) review was performed for the finished device 00001994 number, and no internal action related to the complaint were found during the review.Based on the dhr, the h 4.Device manufacture date has been updated.The issue reported by the customer was confirmed.The odp contains the following caution: always insert a dilator straight into the center of the sheath¿s valve to prevent damage to the valve.Do not insert a dilator at an angle, as damage to the sheath valve may occur.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.This product issue will be addressed through bwi's quality system.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: pc-001190776 additional information was received on (b)(6)2022.This information was inadvertently omitted on follow up report #1( mwr-20102022-0001282982) it was reported that the sheath was being used on the patient.Air did not enter the patient¿s body, therefore were no patient consequences and no treatment was required.No blood return was observed.
 
Manufacturer Narrative
The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (IRVINE)
33 technology drive
irvine CA 92618
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-798
MDR Report Key15490434
MDR Text Key306346852
Report Number2029046-2022-02319
Device Sequence Number1
Product Code DYB
UDI-Device Identifier10846835016253
UDI-Public10846835016253
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K170997
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/09/2023
Device Model NumberD138501
Device Catalogue NumberD138501
Device Lot Number00001994
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received11/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK_CARTO VIZIGO SHEATH
-
-