SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA 500 Back to Search Results
Model Number DIMENSION VISTA 500
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
A falsely depressed carbon dioxide (co2) result was obtained on a patient sample on a dimension vista 500 instrument.The discordant result was not reported to the physician(s).The sample was repeated on an alternate dimension vista 500 instrument.The repeat result was higher than the discordant result.The repeat result was reported, as the correct result, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely depressed co2 result.
 
Additional Manufacturer Narrative
A united states customer contacted a siemens customer care center (ccc) to report that a falsely depressed carbon dioxide (co2) result was obtained on a patient sample on a dimension vista 500 instrument.Siemens reviewed instrument data and determined an issue with the sample mix.A siemens customer service engineer (cse) was dispatched to the customer's site.During the visit, the cse ran the sample 1 (s1) mix test in diagnostic mode and the test failed.The cse replaced the s1 mixer assembly, auto-aligned the s1 arm assembly, and primed the s1 mixer.A quick check was performed which passed.Additionally, quality controls (qc) were performed, which recovered acceptably.Siemens further investigated the issue and determined the potential cause of the falsely depressed co2 result is due to a malfunctioning s1 mixer.The instrument is performing according to specifications.No further evaluation of this device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA 500
Common Device Name
DIMENSION VISTA 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC
101 silvermine road
registration number: 1226181
brookfield CT 06804
Manufacturer Contact
kristin kehyaian
511 benedict ave
tarrytown, NY 10591
9148267101
MDR Report Key15492914
Report Number2517506-2022-00236
Device Sequence Number13952067
Product Code JJE
UDI-Device Identifier00630414593517
UDI-Public00630414593517
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K051087
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDIMENSION VISTA 500
Device Catalogue Number10488217
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/14/2022
Initial Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/30/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-